Autoclave procedure

The delivery of sterile products for use in patient care depends not only on the effectiveness of the sterilization process but also on the unit design, decontamination, disassembling and packaging of the device, loading the sterilizer, monitoring, sterilant quality and quantity, and the appropriateness of the cycle for the load contents, and other aspects of device reprocessing. Healthcare personnel should perform most cleaning, disinfecting, and sterilizing of patient-care supplies in a central processing department in order to more easily control quality.

The aim of central processing is the orderly processing of medical and surgical instruments to protect patients from infections while minimizing risks to staff and preserving the value of the items being reprocessed Healthcare facilities should promote the same level of efficiency and safety in the preparation of supplies in other areas e. Ensuring consistency of sterilization practices requires a comprehensive program that ensures operator competence and proper methods of cleaning and wrapping instruments, loading the sterilizer, operating the sterilizer, and monitoring of the entire process.

Furthermore, care must be consistent from an infection prevention standpoint in all patient-care settings, such as hospital and outpatient facilities. A sterilization process should be verified before it is put into use in healthcare settings. All steam, ETO, and other low-temperature sterilizers are tested with biological and chemical indicators upon installation, when the sterilizer is relocated, redesigned, after major repair and after a sterilization failure has occurred to ensure they are functioning prior to placing them into routine use.

Three consecutive empty steam cycles are run with a biological and chemical indicator in an appropriate test package or tray. Each type of steam cycle used for sterilization e. In a prevacuum steam sterilizer three consecutive empty cycles are also run with a Bowie-Dick test. The sterilizer is not put back into use until all biological indicators are negative and chemical indicators show a correct end-point response, Biological and chemical indicator testing is also done for ongoing quality assurance testing of representative samples of actual products being sterilized and product testing when major changes are made in packaging, wraps, or load configuration.

Biological and chemical indicators are placed in products, which are processed in a full load. When three consecutive cycles show negative biological indicators and chemical indicators with a correct end point response, you can put the change made into routine use Items processed during the three evaluation cycles should be quarantined until the test results are negative.

The central processing area s ideally should be divided into at least three areas: decontamination, packaging, and sterilization and storage. Physical barriers should separate the decontamination area from the other sections to contain contamination on used items.

In the decontamination area reusable contaminatedsupplies and possibly disposable items that are reused are received, sorted, and decontaminated.

The recommended airflow pattern should contain contaminates within the decontamination area and minimize the flow of contaminates to the clean areas. The packaging area is for inspecting, assembling, and packaging clean, but not sterile, material.

Ceilings and wall surfaces should be constructed of non-shedding materials. Physical arrangements of processing areas are presented schematically in four references, Cleaning reduces the bioburden and removes foreign material i. Surgical instruments are generally presoaked or prerinsed to prevent drying of blood and tissue. Precleaning in patient-care areas may be needed on items that are heavily soiled with feces, sputum, blood, or other material. Items sent to central processing without removing gross soil may be difficult to clean because of dried secretions and excretions.

Cleaning and decontamination should be done as soon as possible after items have been used. Several types of mechanical cleaning machines e. This equipment often is automated and may increase productivity, improve cleaning effectiveness, and decrease worker exposure to blood and body fluids.

Delicate and intricate objects and heat- or moisture-sensitive articles may require careful cleaning by hand. All used items sent to the central processing area should be considered contaminated unless decontaminated in the area of originhandled with gloves forceps or tongs are sometimes needed to avoid exposure to sharpsand decontaminated by one of the aforementioned methods to render them safer to handle.

Items composed of more than one removable part should be disassembled.Jump to navigation. Before using the autoclave, check inside for any items left by the previous user that could pose a hazard. Clean the drain strainer before loading the autoclave. Always place items in a secondary container. Do not overload or package bags too tightly. Leave sufficient room for steam circulation. If necessary, place container on its side to maximize steam penetration and avoid entrapment of air.

Use only autoclavable bags to package waste. Do not allow bags to touch the interior walls of the autoclave to avoid melting of plastic. Ensure sufficient liquid is packed with contents of autoclave bags if dry. Place soiled glassware and lab ware in secondary containers and autoclave them in the solids cycle.

autoclave procedure

Loosen caps or use vented closures. In case of clean glassware and wrapped instruments, lay them in a secondary container before autoclaving in wrapped goods cycle. For secondary containment, use autoclave trays made out of polypropylene, polycarbonate or stainless steel. The trays should have a solid bottom and sides to contain the contents and catch spills. Choose appropriate cycle for the material.

Incorrect selection of cycle may damage the autoclave, cause liquid to boil over or bottles to break. Start your cycle and fill out the autoclave user log. A completed cycle usually takes between 1 to 1. Close and lock door. Do not attempt to open the door while autoclave is operating. Unloading Ensure cycle has completed and both temperature and pressure have returned to a safe range. Wear PPE described above, plus an apron and face shield if removing liquids.

Stand back from the door as a precaution and carefully open door no more than 1 inch. This will release residual steam and allow pressure within liquids and containers to normalize. Allow the autoclaved load to stand for 10 minutes in the chamber. This will allow steam to clear and trapped air to escape from hot liquids, reducing risk to operator. Do not agitate containers of super-heated liquids or remove caps before unloading.Skip to main content.

Autoclaves use pressurized steam to destroy microorganisms, and are the most dependable systems available for the decontamination of laboratory waste and the sterilization of laboratory glassware, media, and reagents. For efficient heat transfer, steam must flush the air out of the autoclave chamber. Before using the autoclave, check the drain screen at the bottom of the chamber and clean if blocked.

If the sieve is blocked with debris, a layer of air may form at the bottom of the autoclave, preventing efficient operation. Polypropylene bags are impermeable to steam, and for this reason should not be twisted and taped shut, but gathered loosely at the top and secured with a large rubber band or autoclave tape.

This will create an opening through which steam can penetrate. Search form Search. Research Safety. Commonly called biohazard or autoclave bags, these bags are tear resistant, but can be punctured or burst in the autoclave.

Therefore, place bags in a rigid container during autoclaving. Bags are available in a variety of sizes, and some are printed with an indicator that changes color when processed.

Polypropylene containers and pans. Polypropylene is a plastic capable of withstanding autoclaving, but resistant to heat transfer. Therefore, materials contained in a polypropylene pan will take longer to autoclave than the same materials in a stainless steel pan.

To decrease the time required to sterilize material in these containers, remove the lid if applicable. Stainless steel containers and pans. Stainless steel is a good conductor of heat and is less likely to increase sterilizing time, though is more expensive than polypropylene.

Preparation and Loading of Materials Fill liquid containers only half full. Loosen caps or use vented closures. Always put bags of biological waste into pans to catch spills. Position biohazard bags on their sides, with the bag neck taped loosely.

Leave space between items to allow steam circulation. Household dishpans melt in the autoclave. Use autoclavable polypropylene or stainless steel pans. Cycle Selection Use liquid cycle slow exhaust when autoclaving liquids, to prevent contents from boiling over. Select fast exhaust cycle for glassware. Use fast exhaust and dry cycle for wrapped items. Time Selection Take into account the size of the articles to be autoclaved.

A 2-liter flask containing 1 liter of liquid takes longer to sterilize than four mL flasks each containing mL of liquid. Material with a high insulating capacity animal bedding, high sided polypropylene containers increases the time needed for the load to reach sterilizing temperatures.

Autoclave bags containing biological waste should be autoclaved for 50 minutes to assure decontamination. Removing the Load Check that the chamber pressure is zero. Wear lab coat, eye protection, heat insulating gloves, and closed-toe shoes. Stand behind door when opening it. Slowly open door only a crack. Beware of rush of steam. After the slow exhaust cycle, open autoclave door and allow liquids to cool for 20 minutes before removing.The autoclave is a sealed device similar to a pressure cooker that kills microorganisms using saturated steam under pressure.

The use of moist heat facilitates the killing of all microorganisms, including heat-resistant endospores which is achieved by heating the materials inside the device at temperatures above the boiling point of water.

According to the principle of gas laws, this can be achieved by raising the pressure inside the device. The boiling point vapor pressure equals that of the surrounding atmosphere of water varies depending upon the surrounding environmental pressure. So, in an enclosed device, if we raise the pressure, the temperature at which water boils also increases. The usual procedure is to heat at 1. To make sure, sterilization is successful one should ensure:.

Note that it is not the pressure of the autoclave that kills the microorganisms but the high temperature that can be achieved when steam is placed under pressure. Extended times are also required when large volumes of liquids are being autoclaved because large volumes take longer to reach sterilization temperature. Autoclave comprises of three parts: a pressure chamber, a lid and an electrical heater. The lid is fastened by screw clamps and rendered airtight by an asbestos washer.

autoclave procedure

The lid bears the following. An electrical heater is attached to the jacket; that heats the water to produce steam. Modern autoclaves have devices to maintain proper pressure and record internal temperature during operation.

Regardless of the presence of such a device, autoclave pressure should be checked periodically and maintained. Several methods are available to ensure that autoclaving achieve sterility.

The effectiveness of the sterilization done by autoclave can be monitored by:. Autoclave is particularly useful for media containing water that cannot be sterilized by dry heat.

It is the method of choice for sterilizing the following:. I want autocleave wast lab contain a plates, tubes, ect… Can I put these materials in container then autocleave. I wonder if this technique affect the autocleave.

This site uses Akismet to reduce spam. Learn how your comment data is processed. Culture Media Tests microbes Difference Between. News Ticker.

About Nisha Rijal 30 Articles. Key areas of my work lies in Bacteriology, especially in Antimicrobial resistance. Next Plasmids: Properties, Types and Functions. Greeting, I want autocleave wast lab contain a plates, tubes, ect… Can I put these materials in container then autocleave. Zainab Almahdi. Do you have any queries? Please leave me in the comments section below.

I will be happy to read your comments and reply. Cancel reply.Tape indicators are adhesive-backed paper tape with heat sensitive, chemical indicator markings. Tape indicators are typically placed on the exterior of the waste load. If tape indicators fail on two consecutive loads, notify your Department Safety Manager.

Tape indicators are not designed nor intended to prove that organisms have actually been killed. Chemical color change indicators can be placed within the waste load. If the chemical indicators fail on two consecutive loads, notify your Department Safety Manager. Biological indicator vials contain spores from B.

Autoclaves used to treat biological waste will be evaluated with a biological indicator by EHS on a quarterly basis. Autoclave Use Autoclaves provide a physical method for disinfection and sterilization. They work with a combination of steam, pressure and time. Autoclaves operate at high temperature and pressure in order to kill microorganisms and spores.

They are used to decontaminate certain biological waste and sterilize media, instruments and lab ware. Regulated medical waste that might contain bacteria, viruses and other biological material are recommended to be inactivated by autoclaving before disposal.

Increased cycle time may be necessary depending upon the make-up and volume of the load. The rate of exhaust will depend upon the nature of the load. Dry material can be treated in a fast exhaust cycle, while liquids and biological waste require slow exhaust to prevent boiling over of super-heated liquids. Plastic Polypropylene is an inexpensive resin that can resist autoclave temperatures. Polypropylene containers are often used as secondary containers to hold materials that are autoclaved.

Polycarbonate can also withstand high temperatures. Gloves Gloves must be placed inside of an autoclavable biohazard bag and exposed to a steam setting; gloves will melt slightly but will not burn when autoclaved in this manner. Stainless steel Most metals are designed for extreme conditions and are intended to be sterilized. Make sure to remove any plastics, liners and other items that may melt or combust.

Paper Paper is combustible and should not be placed directly inside an autoclave. It should be autoclaved in a waste bag on a biobag setting to prevent fire. Media Solution No liquid should be sealed in a container and autoclaved. They should autoclaved in a steam producing cycle. Pipette tips Most pipette tips are autoclavable. Some of these tips are plastic, some are high density polyethylene.

In general, pipette tips should only enter the autoclave as waste inside of an approved biohazards bag and always sterilized on a steam-producing setting. Before using the autoclave, check inside for any items left by the previous user that could pose a hazard.

Clean the drain strainer before loading the autoclave. Always place items in a secondary container. Do not overload or package bags too tightly.Of all the methods available for sterilization, moist heat in the form of saturated steam under pressure is the most widely used and the most dependable. Like all sterilization processes, steam sterilization has some deleterious effects on some materials, including corrosion and combustion of lubricants associated with dental handpieces ; reduction in ability to transmit light associated with laryngoscopes ; and increased hardening time 5.

Sterilizing Practices

The basic principle of steam sterilization, as accomplished in an autoclave, is to expose each item to direct steam contact at the required temperature and pressure for the specified time. Thus, there are four parameters of steam sterilization: steam, pressure, temperature, and time.

Specific temperatures must be obtained to ensure the microbicidal activity. At constant temperatures, sterilization times vary depending on the type of item e. The two basic types of steam sterilizers autoclaves are the gravity displacement autoclave and the high-speed prevacuum sterilizer.

In the former, steam is admitted at the top or the sides of the sterilizing chamber and, because the steam is lighter than air, forces air out the bottom of the chamber through the drain vent. The gravity displacement autoclaves are primarily used to process laboratory media, water, pharmaceutical products, regulated medical waste, and nonporous articles whose surfaces have direct steam contact.

For gravity displacement sterilizers the penetration time into porous items is prolonged because of incomplete air elimination.

The advantage of using a vacuum pump is that there is nearly instantaneous steam penetration even into porous loads. A commercially available Bowie-Dick-type test sheet should be placed in the center of the pack.

Autoclave: Principle, Procedure, Types and Uses

Air that is not removed from the chamber will interfere with steam contact. Smaller disposable test packs or process challenge devices have been devised to replace the stack of folded surgical towels for testing the efficacy of the vacuum system in a prevacuum sterilizer. Entrapped air will cause a spot to appear on the test sheet, due to the inability of the steam to reach the chemical indicator. If the sterilizer fails the Bowie-Dick test, do not use the sterilizer until it is inspected by the sterilizer maintenance personnel and passes the Bowie-Dick test.

Another design in steam sterilization is a steam flush-pressure pulsing process, which removes air rapidly by repeatedly alternating a steam flush and a pressure pulse above atmospheric pressure. Air is rapidly removed from the load as with the prevacuum sterilizer, but air leaks do not affect this process because the steam in the sterilizing chamber is always above atmospheric pressure.

Like other sterilization systems, the steam cycle is monitored by mechanical, chemical, and biological monitors. Steam sterilizers usually are monitored using a printout or graphically by measuring temperature, the time at the temperature, and pressure. Typically, chemical indicators are affixed to the outside and incorporated into the pack to monitor the temperature or time and temperature.

Portable table-top steam sterilizers are used in outpatient, dental, and rural clinics. The ability of the sterilizer to reach physical parameters necessary to achieve sterilization should be monitored by mechanical, chemical, and biological indicators.

The oldest and most recognized agent for inactivation of microorganisms is heat. Because a D-value can be determined at various temperatures, a subscript is used to designate the exposure temperature i. Moist heat destroys microorganisms by the irreversible coagulation and denaturation of enzymes and structural proteins.Autoclaving is the most effective and reliable means of sterilizing laboratory materials. Due to the use of pressure, steam and high temperatures, there is significant risk for injury, so it's important for individuals to be properly trained on operational procedures.

The reason for autoclaving infectious waste is that it must be handled several times during transport; proper containment and treatment at the source reduces the potential for an accidental exposure.

The necessary treatment to achieve sterility will vary in relation to the volume of material treated, its contamination level, moisture content and other factors. The following are examples of compatible and incompatible materials. This is not an exhaustive list. Important Notice: Autoclaving hazardous materials may generate toxic vapors or explosive environments.

The proper packaging and containment of infectious materials are crucial to achieve effective sterilization. The most frequent reason for sterilization failure is the lack of contact between the steam and microorganisms. Dry material should be separated from liquid material to achieve proper sterilization.

Skip to main content. University of Iowa Search. Autoclave Compatible and Incompatible Material. General Guidelines. Every autoclave and sterilizer should be inspected and serviced on a regular basis. This will help ensure the equipment is functioning properly. Each unit should have a standard operating procedure written in sufficient detail to ensure that operators will use the equipment properly; controls vary between brands, with each having unique loading characteristics, load-sizing requirements, and cycle setting and types.

Units should be tested regularly with a commercial preparation containing Geobacillus stearothermophilus spores a biological indicatorin particular, any unit in a BSL3 facility. Tape indicators autoclave tape with heat sensitive, chemical indicators should be used in every autoclave load. Note: the indicators only verify that the autoclave has reached normal operating temperatures; they do not indicate that the contents were heated for the appropriate length of time or at the proper pressure.

Therefore, tape indicators cannot be used to prove organisms are actually killed during an autoclave run. Keep detailed records on biological tests, recording thermometers, and service work performed on the unit. High density wastes or materials that insulate the agents from heat and steam penetration are not suitable for steam sterilization. Items that are covered with dirt or film require additional retention times. The importance of properly cleaning items to be sterilized cannot be over emphasized.

autoclave procedure

Place all autoclaved infectious waste into red biohazard bags for disposal. An online training video was developed by Arizona State University that offers safety information, examples of waste to be autoclaved, and procedures for spore testing that may be helpful to any user.

Steam Sterilization

Most bags that are marketed as autoclavable are not suitable if closed because the steam will not penetrate them. Steam resistant bags must be left open or have holes punched into the top to allow the steam to penetrate.

Do not transfer open bags to the autoclave. Never close autoclave bags that have a printed warning stating they are to remain open during sterilization. If air remains trapped in the bag, the material may not be properly sterilized.

Autoclave bags that allow steam penetration tend to melt or crumble during the sterilization process; autoclavable bags can also leak so they should be placed into a shallow stainless steel pan. Plastic pans are less effective because they do not transfer heat as fast or efficiently. Liquid material To prevent bottles from shattering during pressurization and to facilitate steam penetration, bottle caps and stoppers must be loosened after placement into the chamber.


thoughts on “Autoclave procedure”